操作HPLC儀器分析中,如果移動相已吸附有相當含量的O2, N2, CO2等溶存氣體,當這種含有溶存氣移動相被Pump抽送時,Pump會受到溶存氣積成氣泡影響,使得Pump供輸量不足,壓力不穩,難有安定基線,甚至進行兩液移動相混合的梯度 ( Gradient ) 操作,混合比例失真無法作好梯度分析操作。為預防此等狀況的發生,建議選用 “移動相線上真空脫氣設備” (Vacuum on-line degasser),即移動相進入Pump前,先經 “高度真空環境下卓越透氣性的 Teflon AF管”,移動相溶存氣於真空吸引作用下,將會被脫氣,成為沒有溶存氣,純淨的移動相,繼之被Pump供輸進行安穩HPLC分析工作。
(1) NEW BIOTECH ON-LINE DEGASSER
(2) Overall Dimensions
(3) Dissolved gases are actively removed from a flowing liquid stream by vacuum via the Systec AF membrane.
(4) Typical Degasser Implementation

型號 適用流量
P/N 0001-6352A
2-Channel Degasser 2003, 480ul chamber
2ml
P/N 0001-6353A
3-Channel Degasser 2003, 480ul chamber
2ml
P/N 0001-6354A
4-Channel Degasser 2003, 480ul chamber
2ml
P/N 0001-6622 GPC Compatible
2-channel Degasser 2003, 480ul chamber
2ml
P/N 0001-6623 GPC Compatible
3-channel Degasser 2003, 480ul chamber
2ml
P/N 0001-6624 GPC Compatible
4-channel Degasser 2003, 480ul chamber
2ml
型號 適用流量
P/N 0001-6481
2-Channel Degasser 2003, 4.2ml chamber
15ml/min
P/N 0001-6482
2-Channel Degasser 2003, 8.4ml chamber
32ml/min
P/N 0001-6483
2-Channel Degasser 2003, 11.6ml chamber
44ml/min